NE-1000
Single-Channel Infusion Pump
Single-Channel Infusion Pump






















































NE-1000 is a single channel syringe pump designed for automation. User can create multi-phase programs, and the value of every parameter (e.g. pumping rate, dispensing volume,...etc.) can be set individually for each phase. NE-1000 can hold plastic, glass or stainless steel syringes ranging from microliter to 140 mL in volume. The system can provide infusion rates from 0.727 μL/hr (using 1 mL syringe) to 2120 mL/hr (using 60 mL syringe). Lower (‹0.727 μL/hr) and higher (›2120 mL/hr) infusion rates are possible with smaller (‹1 mL) and bigger (›60 mL) syringes, respectively.
| Operating Mode | Infusion, Withdrawal |
|---|---|
| Number of Syringe Channels | 1 |
| Syringe Size | Up to 140 mL 140 mL syringes can be partially filled. 60 mL or smaller syringes can be fully filled. |
| Syringe Inside Diameter (ID) Range | 0.1-50 mm |
| Maximum Pumping Rate | B-D 60 mL syringe: 1699 mL/hr Terumo 60 mL syringe: 2120 mL/hr Maximum pumping rate varies between brands due to differences in syringe ID. |
| Minimum Pumping Rate | B-D 1 mL syringe: 0.73 μL/hr Monoject 1 mL syringe: 1.088 μL/hr Minimum pumping rate varies between brands due to differences in syringe ID. |
| Dispensing Accuracy | ±1% |
| Maximum Force | 35 lbs at minimum speed, 18 lbs at maximum speed |
| Number of Program Phases | 41 |
| RS-232 Pump Network | 100 pumps maximum |
| RS-232 Selectable Baud Rates | 300, 1200, 2400, 9600, 19200 |
| Dimensions | 8.75 in x 5.75 in x 4.5 in (22.86 cm x 14.605 cm x 11.43 cm) |
| Weight | 3.6 lbs (1.63 kg) |
| Voltage | 12V DC at full load |
| Amperage | 750 mA at full load |